Innovation Electronic Co.,Ltd
Use our Machine To Make Your Weighing and Packing Line Easier, Faster and More Cost Saving!
Product Details:
Payment & Shipping Terms:
|
Type: | Single Point Load Cell | Capacity: | 7.2kg To 72kg |
---|---|---|---|
Output: | 2mv/v | Material: | Aluminum |
Application: | Static Weighing | Interchangeable Brand: | HBM |
IP Protection Standard: | IP67 | Accuracy Grade: | C3 |
Hysteresis Error: | 0.0166 | Non-linearity: | 0.0166 |
Highlight: | 7.2kg digital scale load cell,72kg digital scale load cell,Dynamic Weighing single point load cell |
− Accuracy class C3 with OIML-R60 test report
− Max. capacities: 7.2 kg ... 72 kg
− Off center load compensated(OIML R76)
− Degree of protection IP67(according to EN 60 529)
− Shielded connection cable
− optimized for dynamic weighing applications
− Connection cable in six wire circuit
− Different cable lengths
− Aligned output, suitable for
− Aligned output, suitable for connectionin parallel
Type | IPW2D | ||||||
Accuracy class1) | C3; C3MR | ||||||
Number of load cell intervals (nLC) | 3000 | ||||||
Maximum capacity (Emax)2)) | kg | 7.2 | 12 | 18 | 36 | 72 | |
Minimum LC verification interval (vmin) Accuracy class C3 |
g | 1 | 2 | 5 | 10 | 20 | |
Temperature effect on zero balance (TK0) Accuracy class C3
|
% v. Cn / 10 K
|
±0.0194 | ±0.0233 | ±0.0389 | ±0.0389 | ±0.0389 | |
Minimum LC verification interval (vmin) Accuracy class C3MR |
g | 0.5 | 1 | 2 | 5 | 10 | |
Temperature effect on zero balance (TK0) Accuracy class C3MR |
% v. Cn / 10 K
|
±0.0097 | ±0.0116 | ±0.0155 | ±0.00194 | ±0.0194 | |
Max. platform size | mm | 380 x 380 | |||||
Sensitivity (Cn) | mV/V | 2.0±0.2 | |||||
Zero balance | 0±0.1 | ||||||
Temperature effect on sensitivity (TKC)3) Temperature range +20 ... +40℃ [+68 ... +104℉] −10 ... +20℃ [+14 ... +68℉] |
% v. Cn / 10 K
|
±0.0175 ±0.0117 |
|||||
Hysteresis error (dhy)3) | % of Cn | ±0.0166 | |||||
Non-linearity (dlin)3) | ±0.0166 | ||||||
Minimum dead load output return (DR) | ±0.0166 | ||||||
Off center load error4) | ±0.0233 | ||||||
Input resistance (RLC) | Ω | 300 ... 500 | |||||
Output resistance (R0) | 300 ... 500 | ||||||
Reference excitation voltage (Uref) | V | 5 | |||||
Nominal range of excitation voltage (BU) | 1 ... 12 | ||||||
Insulation resistance (Ris) at 100 VDC | GΩ | > 2 | |||||
Nominal temp. range (BT) |
°C [°F] |
−10 ... +40 [+14 ... +104] | |||||
Service temp. range (Btu) | −10 ... +50 [+14 ... +122] | ||||||
Storage temp. range (Btl) | −25 ... +70 [−13 ... +158] | ||||||
Safe load limit (EL ), at max. 100mm eccentricity |
% of Emax |
150 160 |
|||||
Lateral load limit (Elq), static |
% of Emax |
300 | |||||
Breaking load (Ed) | 300 | ||||||
Weight (G), approx. | kg | 0.25 | |||||
Protection class according to EN 60 529 (IEC 529) |
IP67 | ||||||
Material: Measuring element Coating Cable sheath |
Aluminum Silicone rubber PVC |
Contact Person: Miss. TINA LIU
Tel: +8618758163948
Fax: 86-571-8838-5175